# Quantitative Determination of Volatile Organic Pollutants in Water using Headspace–Trap GC–MS

![Markes-volatile-organic-pollutants](https://appnote.sepscience.com/hs-fs/hubfs/Featured_Applications/Oct_18/Markes-volatile-organic-pollutants.png?width=300&name=Markes-volatile-organic-pollutants.png)

## Download your copy...

This study shows that headspace samples acquired on the new Centri automated multi-mode sampling and concentration system for gas chromatography–mass  
spectrometry (GC–MS) can be used to quantify a wide range of volatile organic pollutants in water.

Analysis of an 83-component standard based on the list in US EPA Method 524.2 showed chromatographic performance within the requirements of the method, as indicated by the mean values for linearity (R2 0.9990), reproducibility (6.9% RSD) and analyte recovery (90%), and a mean detection limit of 0.11 ppb.

To download your copy of this application note complete the form opposite.

Registrants for this application note are covered by the Eclipse Business Media [Privacy Policy](http://www.sepscience.com/privacy-policy).

Eclipse Business Media Ltd, Regd in England, No. 06513189, Woodview, Bull Lane Industrial Estate, Sudbury, CO10 0FD, United Kingdom, T +44 (0)161 818 7434, [info@sepscience.com](mailto:info@sepscience.com)

 

Eclipse Business Media Ltd needs the contact information you provide to us to contact you about our products and services. You may unsubscribe from these communications at any time. For information on how to unsubscribe, as well as our privacy practices and commitment to protecting your privacy, check out our [Privacy Policy](https://www.sepscience.com/privacy-policy/) or contact [privacy@labxmediagroup.com](mailto:privacy@labxmediagroup.com) 

Copyright © 2025. All rights reserved

[Contact Us](https://www.sepscience.com/contact-us/)  
[Advertise with Us](https://www.sepscience.com/advertise-with-us/)